top of page

Troubleshooting AMX

Error Number
After stored now behaves abnormally
Error Code / System Message
​
Error Question
What should I do if an XR200amx or XR220amx has been stored and now behaves abnormally?
Short Answer
A system stored at low temperature or with batteries connected/discharged can show low-temperature, low-charge, shutdown and generator errors without necessarily requiring parts.
Possible Causes
Cold system; self-discharge; very low state of charge; improper storage.
Troubleshooting Steps
Allow a cold system to stabilize to the specified environment. Fully recharge it using the approved charging practice, then reassess errors before replacing batteries or generator components.
Parts Areas to Check
​
For Related Models
Optima XR200amx, Optima XR220amx