top of page
Troubleshooting AMX
Error Number
FlashPad grid issues
Error Code / System Message
Error Question
Which diagnostic checks the FlashPad grid on XR220amx?
Short Answer
The Grid Sensor diagnostic verifies detector recognition of an attached grid or tests the user interaction/status display depending on operating mode.
Possible Causes
Grid attachment/orientation/sensor issue.
Troubleshooting Steps
Run Grid Sensor and confirm the detector recognizes the expected grid state. Swap with a known-good grid when available to isolate the accessory from the detector.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx