Short Answer
Generator tests include rotation, rotor impedance, power supplies, resonant components, inverter, 70C, rise-time/regulation, filament-driver, mA, kV and Auto tests.
Possible Causes
Generator, tube, filament, rotor, resonant, inverter, DC-bus, mA or kV subsystem fault.
Troubleshooting Steps
Match the error-log subsystem to the diagnostic. Use compound kV, mA or Auto tests when appropriate and follow the service-manual sequence for any failed subtest.
Parts Areas to Check