top of page

Troubleshooting AMX

Error Number
System power problem
Error Code / System Message
​
Error Question
Which XR240amx diagnostics should I run for a system power problem?
Short Answer
System Power diagnostics include Power Control Comm, Power Control Limits, Power Supply Thermal, Collimator Light and other power-state tests.
Possible Causes
Power-control communication; battery voltage/current; thermal issue; AC/breaker status.
Troubleshooting Steps
Open SUIF Diagnostics and select System Power. Run the test that matches the symptom, then follow the on-screen instructions and compare actual values with limits.
Parts Areas to Check
​
For Related Models
Optima XR240amx
bottom of page