top of page
Troubleshooting AMX
Error Number
Detector charging lights are not illuminated
Error Code / System Message
Error Question
Why are the detector charging lights on my XR240amx bin not illuminated?
Short Answer
A dark charging indicator does not always mean the detector is not charging; the UI and detector logs should be checked.
Possible Causes
Defective LED membrane or indicator cable; BCB fault; 15 VDC loss.
Troubleshooting Steps
Verify the detector is recognized in the UI and check the detector battery log for increasing charge. If charge is increasing but the bin LEDs remain off, inspect the LED membrane and indicator cable. If charging itself is also absent, continue BCB/power troubleshooting.
Parts Areas to Check
For Related Models
Optima XR240amx
bottom of page