top of page

Troubleshooting AMX

Error Number
Digital power supply does not function
Error Code / System Message
​
Error Question
Why can the Definium AMX 700 Digital Detector Power Supply fail after fuse replacement?
Short Answer
The manual states the fuses must be installed in the correct orientation or the power supply may not function.
Possible Causes
Fuse installed in the wrong orientation; blown fuse; power-supply fault.
Troubleshooting Steps
Verify each Digital Detector Power Supply fuse is installed in the orientation shown in the manual. If power is still absent, continue with debug-header voltage and fuse checks.
Parts Areas to Check
Digital Detector Power Supply; 2A/3A/5A fuses
For Related Models