top of page
Troubleshooting AMX
Error Number
Booted and then shut down or trip the breaker
Error Code / System Message
Error Question
Why does my XR240amx boot and then shut down or trip the breaker?
Short Answer
The system-battery troubleshooting section identifies very low battery state of charge as a condition that can allow a short boot before the breaker trips.
Possible Causes
Low system battery state of charge; low pack voltage.
Troubleshooting Steps
With qualified service access, record individual battery voltages and current using Power Control Limits diagnostics. Fully charge the system uninterrupted for 4–6 hours, then retest unplugged and remove surface charge before reassessing.
Parts Areas to Check
For Related Models
Optima XR240amx
bottom of page