top of page
Troubleshooting AMX
Error Number
Detector warn that the detector grid is backward
Error Code / System Message
Error Question
Why does the XR240amx warn that the detector grid is backward?
Short Answer
The detector automatically senses grid orientation and warns when the grid is attached on the wrong side/orientation.
Possible Causes
Grid installed backward or on wrong side of detector.
Troubleshooting Steps
Remove the grid and reinstall it in the correct orientation and attachment position. Confirm the warning clears before imaging.
Parts Areas to Check
For Related Models
Optima XR240amx