top of page

Troubleshooting AMX

Error Number
Detector not charging in the bin
Error Code / System Message
​
Error Question
Why is my GE Optima XR240amx detector not charging in the bin?
Short Answer
The manual lists the detector battery, charging cable, docking connector, bin control board, LVLE power, and indicator circuitry as possible causes.
Possible Causes
Defective battery; damaged charging cable; stuck/faulty docking connector; BCB failure; loss of 15 VDC or 24 VDC; indicator membrane/cable fault.
Troubleshooting Steps
Confirm the detector is recognized. Inspect the battery, charging cable and spring-loaded docking connector. Verify the detector is in the correct slot. If charging is shown in the UI/logs but LEDs are dark, isolate the LED membrane/indicator cable. Qualified service personnel can continue electrical checks.
Parts Areas to Check
​
For Related Models
Optima XR240amx