top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

DRIVE BYTE LIMIT

Error Code System Message

DRIVE BYTE LIMIT - Drive byte calibration limit exceeded

DRIVE BYTE LIMIT, Drive byte calibration limit exceeded, AMX-4 DRIVE BYTE LIMIT, AMX-4+ DRIVE BYTE LIMIT, Definium AMX 700 DRIVE BYTE LIMIT, GE AMX error DRIVE BYTE LIMIT
Question

What does DRIVE BYTE LIMIT mean on a GE AMX-Series mobile X-ray?

Short Answer

DRIVE BYTE LIMIT: Drive byte calibration limit exceeded.

Possible Causes

Drive upper/lower calibration limit exceeded.

Troubleshooting Steps

Check drive calibration/current feedback and recalibrate drive.

Parts Areas to Check

CPU board; calibration data; subsystem identified by the error

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

DRIVE WORD LIMIT

Error Code System Message

DRIVE WORD LIMIT - Drive word calibration limit exceeded

DRIVE WORD LIMIT, Drive word calibration limit exceeded, AMX-4 DRIVE WORD LIMIT, AMX-4+ DRIVE WORD LIMIT, Definium AMX 700 DRIVE WORD LIMIT, GE AMX error DRIVE WORD LIMIT
Question

What does DRIVE WORD LIMIT mean on a GE AMX-Series mobile X-ray?

Short Answer

DRIVE WORD LIMIT: Drive word calibration limit exceeded.

Possible Causes

Drive upper/lower word calibration limit exceeded.

Troubleshooting Steps

Check drive calibration/current feedback and recalibrate drive.

Parts Areas to Check

CPU board; calibration data; subsystem identified by the error

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Display Goes Out On Unit Intermittently

Error Code System Message

​

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0167, Display Goes Out On Unit Intermittently
Question

How do I troubleshoot this AMX-4 Series issue: Display Goes Out On Unit Intermittently?

Short Answer

Check the 3 Pins On J1 (Feeds Battery Voltage To The LVLE Power Supply) And Make Sure They Are Not Burnt

Possible Causes

Problem: Display Goes Out On Unit Intermittently

Troubleshooting Steps

Check the 3 Pins On J1 (Feeds Battery Voltage To The LVLE Power Supply) And Make Sure They Are Not Burnt

Parts Areas to Check

LVLE Power Supply; Battery

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Display issues, trashy, scrolling or no display during charge.

Error Code System Message

​

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0152, Display issues, trashy, scrolling or no display during charge.
Question

How do I troubleshoot this AMX-4 Series issue: Display issues, trashy, scrolling or no display during charge?

Short Answer

Monitor LVLE +5 V with an oscilloscope for about 15 minutes and look for breakup/noise on the +5 V supply. If noise is evident, replace the LVLE. If LVLE appears OK, check the Display kV/mA Charge Bar Board (Kit Part #2139946) or Display Controller Board (Part #46-264982G1). Battery Charger Board failure is also possible; tap on relays to see if display failure occurs. Solution: replace LVLE Power Supply, Display kV/mA Charge Bar Board, Display Controller Board and/or Battery Charger Board as indicated by troubleshooting.

Possible Causes

Problem: Display issues, trashy, scrolling or no display during charge.

Troubleshooting Steps

Monitor LVLE Power Supply +5 volt with a scope for about 15 minutes - look for break-up (noise) on +5 volt supply, If noise is evident, replace the LVLE Power Supply. If LVLE Power Supply appears OK, this could also be caused by a bad Display Board (replace with orderable kit Part # 2139946) or Display Controller Board (Part # 46-264982G1). Also could be possible Battery Charger Board failure. Check tap on relays to see if display failure occurs. Replace LVLE Power Supply, Display kV mA Charge Bar Board, Display Controller Board and/or Battery Charger Boards

Parts Areas to Check

Battery Charger Board; LVLE Power Supply; Display Controller Board; Display Board; Battery

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Drive calibration required

Error Code System Message

HNDL CAL ERR 1-4

AMX drive handle calibration, HNDL CAL ERR
Question

When should I run AMX-Series drive-handle calibration?

Short Answer

Drive-handle calibration is required when the drive handle/transducer range or zero point is incorrect or related HNDL CAL errors occur.

Possible Causes

Handle transducer replacement/misalignment; calibration drift.

Troubleshooting Steps

Enter service calibration mode, perform the Drive Handle Calibration procedure, and confirm zero/forward/reverse values are within range before returning the unit to service.

Parts Areas to Check

Drive handle; Hall-effect transducers; CPU inputs

For Related Models

AMX-4, AMX-4+, Definium AMX 700

bottom of page