AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0201
Error Code System Message
Question
What does error 0201 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0201: Fil large: Filament over current detected.
Possible Causes
Subsystem/category: Filament Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0202
Error Code System Message
Question
What does error 0202 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0202: Fil large: Filament open circuit detected.
Possible Causes
Subsystem/category: Filament Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0203
Error Code System Message
Question
What does error 0203 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0203: Fil large: Filament drive current exceeds max limit.
Possible Causes
Subsystem/category: Filament Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0204
Error Code System Message
Question
What does error 0204 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0204: Fil large: Filament drive current below min limit.
Possible Causes
Subsystem/category: Filament Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0205
Error Code System Message
Question
What does error 0205 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0205: Fil large: Filament inverter is off and current not zero.
Possible Causes
Subsystem/category: Filament Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check