Error 0207

Error Code System Message

​

0207, error 0207, XR200 0207, XR220 0207, Optima XR200amx error 0207, Optima XR220amx error 0207, Fil large: Over current detection on filament
Question

What does error 0207 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0207: Fil large: Over current detection on filament.

Possible Causes

Subsystem/category: Filament Large. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0208

Error Code System Message

​

0208, error 0208, XR200 0208, XR220 0208, Optima XR200amx error 0208, Optima XR220amx error 0208, Fil small: 48V supply filament circuit out of range [44V-52V]
Question

What does error 0208 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0208: Fil small: 48V supply filament circuit out of range [44V-52V].

Possible Causes

Subsystem/category: Filament Large. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0281

Error Code System Message

​

0281, error 0281, XR200 0281, XR220 0281, Optima XR200amx error 0281, Optima XR220amx error 0281, Fil small: Unexpected generator firmware error
Question

What does error 0281 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0281: Fil small: Unexpected generator firmware error.

Possible Causes

Subsystem/category: Filament Large. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0283

Error Code System Message

​

0283, error 0283, XR200 0283, XR220 0283, Optima XR200amx error 0283, Optima XR220amx error 0283, Fil large: Autodiag failed during generator boot
Question

What does error 0283 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0283: Fil large: Autodiag failed during generator boot.

Possible Causes

Subsystem/category: Filament Large. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0401

Error Code System Message

​

0401, error 0401, XR200 0401, XR220 0401, Optima XR200amx error 0401, Optima XR220amx error 0401, Rotor: Lost rotor communication between Power and Main Aux modules
Question

What does error 0401 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0401: Rotor: Lost rotor communication between Power and Main Aux modules.

Possible Causes

Subsystem/category: Rotor. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics; inspect tube-stator/rotor wiring and Power/Main Auxiliary communication.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx