Error Number
Error 0207
Error Code System Message
Question
What does error 0207 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0207: Fil large: Over current detection on filament.
Possible Causes
Subsystem/category: Filament Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0208
Error Code System Message
Question
What does error 0208 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0208: Fil small: 48V supply filament circuit out of range [44V-52V].
Possible Causes
Subsystem/category: Filament Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0281
Error Code System Message
Question
What does error 0281 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0281: Fil small: Unexpected generator firmware error.
Possible Causes
Subsystem/category: Filament Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0283
Error Code System Message
Question
What does error 0283 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0283: Fil large: Autodiag failed during generator boot.
Possible Causes
Subsystem/category: Filament Large. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0401
Error Code System Message
Question
What does error 0401 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0401: Rotor: Lost rotor communication between Power and Main Aux modules.
Possible Causes
Subsystem/category: Rotor. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics; inspect tube-stator/rotor wiring and Power/Main Auxiliary communication.