top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0488

Error Code System Message

​

0488, error 0488, XR200 0488, XR220 0488, Optima XR200amx error 0488, Optima XR220amx error 0488, Rotor: Speed communication failure between Power and Main Aux modules
Question

What does error 0488 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0488: Rotor: Speed communication failure between Power and Main Aux modules.

Possible Causes

Subsystem/category: Rotor. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics; inspect tube-stator/rotor wiring and Power/Main Auxiliary communication.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0489

Error Code System Message

​

0489, error 0489, XR200 0489, XR220 0489, Optima XR200amx error 0489, Optima XR220amx error 0489, Rotor: Generator detected incompatible rotor board circuit
Question

What does error 0489 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0489: Rotor: Generator detected incompatible rotor board circuit.

Possible Causes

Subsystem/category: Rotor. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics; inspect tube-stator/rotor wiring and Power/Main Auxiliary communication.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0500

Error Code System Message

​

0500, error 0500, XR200 0500, XR220 0500, Optima XR200amx error 0500, Optima XR220amx error 0500, Input stage: Unexpected generator firmware error
Question

What does error 0500 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0500: Input stage: Unexpected generator firmware error.

Possible Causes

Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: ABORT.

Troubleshooting Steps

Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0501

Error Code System Message

​

0501, error 0501, XR200 0501, XR220 0501, Optima XR200amx error 0501, Optima XR220amx error 0501, Input stage: DC Bus command out of range [100V-420V]
Question

What does error 0501 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0501: Input stage: DC Bus command out of range [100V-420V].

Possible Causes

Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: ABORT.

Troubleshooting Steps

Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0503

Error Code System Message

​

0503, error 0503, XR200 0503, XR220 0503, Optima XR200amx error 0503, Optima XR220amx error 0503, Input stage: DC Bus rise time too fast during DC Bus charge [DC Bus
Question

What does error 0503 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0503: Input stage: DC Bus rise time too fast during DC Bus charge [DC Bus < 200V].

Possible Causes

Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: ABORT.

Troubleshooting Steps

Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page