AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0523
Error Code System Message
Question
What does error 0523 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0523: Input stage: Input stage damaged.
Possible Causes
Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0524
Error Code System Message
Question
What does error 0524 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0524: Input stage: Hardware over input current detected [Iin > 140A].
Possible Causes
Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0525
Error Code System Message
Question
What does error 0525 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0525: Input stage: Boot timeout when boot is started.
Possible Causes
Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0580
Error Code System Message
Question
What does error 0580 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0580: Input stage: DC Bus timeout when boot is started.
Possible Causes
Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: ABORT.
Troubleshooting Steps
Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0581
Error Code System Message
Question
What does error 0581 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0581: Input stage: Low input voltage during boost while Vin < 95V.
Possible Causes
Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.
Parts Areas to Check