AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0606
Error Code System Message
Question
What does error 0606 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0606: kV: Too many spits detected [spits count > 8 per second].
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics and follow the specific code workflow.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0607
Error Code System Message
Question
What does error 0607 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0607: kV: Inverter current mismatch between measured & expected values..
Possible Causes
Subsystem/category: kV. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics and follow the specific code workflow.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0608
Error Code System Message
Question
What does error 0608 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0608: kV: Inverter over current detected [ILR > 250A].
Possible Causes
Subsystem/category: kV. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics and follow the specific code workflow.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0609
Error Code System Message
Question
What does error 0609 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0609: kV: Measured kV is greater than target before exposure start.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics and follow the specific code workflow.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 060A
Error Code System Message
Question
What does error 060A mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 060A: kV: Measured kV too low after exposure start [kV < 3kV].
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics and follow the specific code workflow.
Parts Areas to Check