AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0616
Error Code System Message
Question
What does error 0616 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0616: kV: Techniques demanded cannot be achieved.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics and follow the specific code workflow.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0681
Error Code System Message
Question
What does error 0681 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0681: kV: Synchronization issue at exposure start.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics and follow the specific code workflow.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0684
Error Code System Message
Question
What does error 0684 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0684: kV: kV detected while inverter is off [kV > 1kV].
Possible Causes
Subsystem/category: kV. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics and follow the specific code workflow.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0685
Error Code System Message
Question
What does error 0685 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0685: kV: Resonant current detected while inverter is off [ILR > 1A].
Possible Causes
Subsystem/category: kV. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics and follow the specific code workflow.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 0701
Error Code System Message
Question
What does error 0701 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0701: Thermal: Tube overtemp switch open.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Allow cooling when appropriate, verify airflow, and run generator thermal/70C diagnostics. Check the board or sensor identified by the code.
Parts Areas to Check