top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 090B

Error Code System Message

​

090B, error 090B, XR200 090B, XR220 090B, Optima XR200amx error 090B, Optima XR220amx error 090B, RTL: X-ray hardware command risen at generator boot up
Question

What does error 090B mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 090B: RTL: X-ray hardware command risen at generator boot up.

Possible Causes

Subsystem/category: Boot. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review generator software/boot state and related errors; attempt approved software recovery/download and reboots before replacing hardware.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 090C

Error Code System Message

​

090C, error 090C, XR200 090C, XR220 090C, Optima XR200amx error 090C, Optima XR220amx error 090C, WDT: Watchdog SW error
Question

What does error 090C mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 090C: WDT: Watchdog SW error.

Possible Causes

Subsystem/category: Boot. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Review generator software/boot state and related errors; attempt approved software recovery/download and reboots before replacing hardware.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0981

Error Code System Message

​

0981, error 0981, XR200 0981, XR220 0981, Optima XR200amx error 0981, Optima XR220amx error 0981, Boot: Autodiag failure detected on diag
Question

What does error 0981 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0981: Boot: Autodiag failure detected on diag.

Possible Causes

Subsystem/category: Boot. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review generator software/boot state and related errors; attempt approved software recovery/download and reboots before replacing hardware.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0A81

Error Code System Message

​

0A81, error 0A81, XR200 0A81, XR220 0A81, Optima XR200amx error 0A81, Optima XR220amx error 0A81, Misc: Internal power supplies out of spec
Question

What does error 0A81 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0A81: Misc: Internal power supplies out of spec.

Possible Causes

Subsystem/category: Misc / Power / Cable. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review the Class M error log for related errors, then run the SUIF diagnostic matching the affected generator subsystem.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0A82

Error Code System Message

​

0A82, error 0A82, XR200 0A82, XR220 0A82, Optima XR200amx error 0A82, Optima XR220amx error 0A82, Cable: Power module's internal cabling is missing or damaged
Question

What does error 0A82 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0A82: Cable: Power module's internal cabling is missing or damaged.

Possible Causes

Subsystem/category: Misc / Power / Cable. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review the Class M error log for related errors, then run the SUIF diagnostic matching the affected generator subsystem.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page