top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0AA0

Error Code System Message

​

0AA0, error 0AA0, XR200 0AA0, XR220 0AA0, Optima XR200amx error 0AA0, Optima XR220amx error 0AA0, Misc: Reboot caused by an unexpected crash
Question

What does error 0AA0 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0AA0: Misc: Reboot caused by an unexpected crash.

Possible Causes

Subsystem/category: Misc / Power / Cable. Manual recovery/error class: DEBUG.

Troubleshooting Steps

Review the Class M error log for related errors, then run the SUIF diagnostic matching the affected generator subsystem.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0AA1

Error Code System Message

​

0AA1, error 0AA1, XR200 0AA1, XR220 0AA1, Optima XR200amx error 0AA1, Optima XR220amx error 0AA1, WDT: Watchdog SW error
Question

What does error 0AA1 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0AA1: WDT: Watchdog SW error.

Possible Causes

Subsystem/category: Misc / Power / Cable. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review the Class M error log for related errors, then run the SUIF diagnostic matching the affected generator subsystem.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0B81

Error Code System Message

​

0B81, error 0B81, XR200 0B81, XR220 0B81, Optima XR200amx error 0B81, Optima XR220amx error 0B81, mA: mA measured while inverter is off [mA > 1mA]
Question

What does error 0B81 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0B81: mA: mA measured while inverter is off [mA > 1mA].

Possible Causes

Subsystem/category: mA Generation. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0B82

Error Code System Message

​

0B82, error 0B82, XR200 0B82, XR220 0B82, Optima XR200amx error 0B82, Optima XR220amx error 0B82, mA: No mAs measured after exposure start [mAs
Question

What does error 0B82 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0B82: mA: No mAs measured after exposure start [mAs < 0.005].

Possible Causes

Subsystem/category: mA Generation. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 0B83

Error Code System Message

​

0B83, error 0B83, XR200 0B83, XR220 0B83, Optima XR200amx error 0B83, Optima XR220amx error 0B83, mA: Inconsistency between mA anode and mA cathode measurement
Question

What does error 0B83 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 0B83: mA: Inconsistency between mA anode and mA cathode measurement.

Possible Causes

Subsystem/category: mA Generation. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page