AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 3606001
Error Code System Message
Question
What does error 3606001 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 3606001: Generator Error - generator in error state.
Possible Causes
Subsystem/category: Generator. Manual recovery/error class: ERROR.
Troubleshooting Steps
Generic generator-in-error-state message; review detailed generator errors and battery-related debug text.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8110
Error Code System Message
Question
What does error 8110 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8110: CANOpen: Communication overflow detected.
Possible Causes
Subsystem/category: CANopen. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review generator communication errors and inspect CANOpen communication/cabling before hardware replacement.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8130
Error Code System Message
Question
What does error 8130 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8130: CANOpen: Communication watchdog failure detected.
Possible Causes
Subsystem/category: CANopen. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review generator communication errors and inspect CANOpen communication/cabling before hardware replacement.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8300101
Error Code System Message
Question
What does error 8300101 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8300101: Logic power out of specification.
Possible Causes
Subsystem/category: System Power. Manual recovery/error class: ERROR.
Troubleshooting Steps
Low logic power can accompany battery problems; evaluate battery state and power distribution before replacing boards.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8300102
Error Code System Message
Question
What does error 8300102 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8300102: One or more Locust board fuses broken.
Possible Causes
Subsystem/category: System Power. Manual recovery/error class: ERROR.
Troubleshooting Steps
Inspect Locust fuses and related low-logic-power conditions; when combined with generator/battery errors, evaluate batteries first.
Parts Areas to Check