top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 3606001

Error Code System Message

​

3606001, error 3606001, XR200 3606001, XR220 3606001, Optima XR200amx error 3606001, Optima XR220amx error 3606001, Generator Error - generator in error state
Question

What does error 3606001 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 3606001: Generator Error - generator in error state.

Possible Causes

Subsystem/category: Generator. Manual recovery/error class: ERROR.

Troubleshooting Steps

Generic generator-in-error-state message; review detailed generator errors and battery-related debug text.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8110

Error Code System Message

​

8110, error 8110, XR200 8110, XR220 8110, Optima XR200amx error 8110, Optima XR220amx error 8110, CANOpen: Communication overflow detected
Question

What does error 8110 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8110: CANOpen: Communication overflow detected.

Possible Causes

Subsystem/category: CANopen. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review generator communication errors and inspect CANOpen communication/cabling before hardware replacement.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8130

Error Code System Message

​

8130, error 8130, XR200 8130, XR220 8130, Optima XR200amx error 8130, Optima XR220amx error 8130, CANOpen: Communication watchdog failure detected
Question

What does error 8130 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8130: CANOpen: Communication watchdog failure detected.

Possible Causes

Subsystem/category: CANopen. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review generator communication errors and inspect CANOpen communication/cabling before hardware replacement.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8300101

Error Code System Message

​

8300101, error 8300101, XR200 8300101, XR220 8300101, Optima XR200amx error 8300101, Optima XR220amx error 8300101, Logic power out of specification
Question

What does error 8300101 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8300101: Logic power out of specification.

Possible Causes

Subsystem/category: System Power. Manual recovery/error class: ERROR.

Troubleshooting Steps

Low logic power can accompany battery problems; evaluate battery state and power distribution before replacing boards.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8300102

Error Code System Message

​

8300102, error 8300102, XR200 8300102, XR220 8300102, Optima XR200amx error 8300102, Optima XR220amx error 8300102, One or more Locust board fuses broken
Question

What does error 8300102 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8300102: One or more Locust board fuses broken.

Possible Causes

Subsystem/category: System Power. Manual recovery/error class: ERROR.

Troubleshooting Steps

Inspect Locust fuses and related low-logic-power conditions; when combined with generator/battery errors, evaluate batteries first.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page