AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 8301001
Error Code System Message
Question
What does error 8301001 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301001: Hardware error – Firefly FPGA temperature too low.
Possible Causes
Subsystem/category: System Power. Manual recovery/error class: ERROR.
Troubleshooting Steps
If the system was stored cold, allow temperature stabilization; otherwise evaluate Firefly temperature sensing and system environment.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301003
Error Code System Message
Question
What does error 8301003 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301003: Hardware error – Firefly Power Supply temperature too low.
Possible Causes
Subsystem/category: System Power. Manual recovery/error class: ERROR.
Troubleshooting Steps
If the system was stored cold, allow temperature stabilization; otherwise evaluate Firefly power-supply temperature sensing.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301043
Error Code System Message
Question
What does error 8301043 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301043: Battery voltage below threshold – trip breaker to protect batteries.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: TRIP.
Troubleshooting Steps
Fully charge the system and review individual pack voltages; repeated recurrence can indicate weak batteries.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301046
Error Code System Message
Question
What does error 8301046 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301046: Battery stack voltage <132V idle or <126V driving.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: TRIP.
Troubleshooting Steps
Fully charge, then reassess under load; repeated recurrence after proper charging suggests aged/weak batteries.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301048
Error Code System Message
Question
What does error 8301048 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301048: Last breaker trip due to battery temperature.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: STATUS.
Troubleshooting Steps
Review the preceding pack-temperature errors and battery thermal condition.
Parts Areas to Check