top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 8301001

Error Code System Message

​

8301001, error 8301001, XR200 8301001, XR220 8301001, Optima XR200amx error 8301001, Optima XR220amx error 8301001, Hardware error – Firefly FPGA temperature too low
Question

What does error 8301001 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301001: Hardware error – Firefly FPGA temperature too low.

Possible Causes

Subsystem/category: System Power. Manual recovery/error class: ERROR.

Troubleshooting Steps

If the system was stored cold, allow temperature stabilization; otherwise evaluate Firefly temperature sensing and system environment.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8301003

Error Code System Message

​

8301003, error 8301003, XR200 8301003, XR220 8301003, Optima XR200amx error 8301003, Optima XR220amx error 8301003, Hardware error – Firefly Power Supply temperature too low
Question

What does error 8301003 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301003: Hardware error – Firefly Power Supply temperature too low.

Possible Causes

Subsystem/category: System Power. Manual recovery/error class: ERROR.

Troubleshooting Steps

If the system was stored cold, allow temperature stabilization; otherwise evaluate Firefly power-supply temperature sensing.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8301043

Error Code System Message

​

8301043, error 8301043, XR200 8301043, XR220 8301043, Optima XR200amx error 8301043, Optima XR220amx error 8301043, Battery voltage below threshold – trip breaker to protect batteries
Question

What does error 8301043 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301043: Battery voltage below threshold – trip breaker to protect batteries.

Possible Causes

Subsystem/category: System Battery. Manual recovery/error class: TRIP.

Troubleshooting Steps

Fully charge the system and review individual pack voltages; repeated recurrence can indicate weak batteries.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8301046

Error Code System Message

​

8301046, error 8301046, XR200 8301046, XR220 8301046, Optima XR200amx error 8301046, Optima XR220amx error 8301046, Battery stack voltage <132V idle or
Question

What does error 8301046 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301046: Battery stack voltage <132V idle or <126V driving.

Possible Causes

Subsystem/category: System Battery. Manual recovery/error class: TRIP.

Troubleshooting Steps

Fully charge, then reassess under load; repeated recurrence after proper charging suggests aged/weak batteries.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8301048

Error Code System Message

​

8301048, error 8301048, XR200 8301048, XR220 8301048, Optima XR200amx error 8301048, Optima XR220amx error 8301048, Last breaker trip due to battery temperature
Question

What does error 8301048 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301048: Last breaker trip due to battery temperature.

Possible Causes

Subsystem/category: System Battery. Manual recovery/error class: STATUS.

Troubleshooting Steps

Review the preceding pack-temperature errors and battery thermal condition.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page