top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 107

Error Code System Message

107

AMX-4, AMX-4+, Definium AMX 700, ERROR 107
Question

What should I check for errors 107 on the AMX-4 Series?

Short Answer

1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.

Possible Causes

Perform recalibration of the “CAL BATT CHARGER” portion. Replace the Battery Charger Board.

Troubleshooting Steps

1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.

Parts Areas to Check

Battery Charger Board; Battery

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 12

Error Code System Message

ERROR 12 - Left drive stall / release-handle fault

ERROR 12, Left drive stall / release-handle fault, AMX-4 ERROR 12, AMX-4+ ERROR 12, Definium AMX 700 ERROR 12, GE AMX error ERROR 12
Question

What does ERROR 12 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 12: Left drive stall / release-handle fault.

Possible Causes

Left stall signal detected.

Troubleshooting Steps

Check motor relays/snubber circuits, brakes, drive motor and stall/release wiring.

Parts Areas to Check

Drive control board; drive handle/Hall sensors; motor relays; drive motors; feedback wiring

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 12

Error Code System Message

12

AMX-4, AMX-4+, Definium AMX 700, ERROR 12, Error 12
Question

What should I check for error 13 on the AMX-4 Series?

Short Answer

Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards

Possible Causes

Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards

Troubleshooting Steps

Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards

Parts Areas to Check

Drive Amplifier Board

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 13

Error Code System Message

13

AMX-4, AMX-4+, Definium AMX 700, ERROR 13, Error 13
Question

What should I check for error 13 on the AMX-4 Series?

Short Answer

Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards

Possible Causes

Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards

Troubleshooting Steps

Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards

Parts Areas to Check

Drive Amplifier Board

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 13

Error Code System Message

ERROR 13 - Right drive stall / release-handle fault

ERROR 13, Right drive stall / release-handle fault, AMX-4 ERROR 13, AMX-4+ ERROR 13, Definium AMX 700 ERROR 13, GE AMX error ERROR 13
Question

What does ERROR 13 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 13: Right drive stall / release-handle fault.

Possible Causes

Right stall signal detected.

Troubleshooting Steps

Check right-side relay/snubber, brakes, drive motor and stall wiring.

Parts Areas to Check

Drive control board; drive handle/Hall sensors; motor relays; drive motors; feedback wiring

For Related Models

AMX-4, AMX-4+, Definium AMX 700

bottom of page