AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 107
Error Code System Message
107
Question
What should I check for errors 107 on the AMX-4 Series?
Short Answer
1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Possible Causes
Perform recalibration of the “CAL BATT CHARGER” portion. Replace the Battery Charger Board.
Troubleshooting Steps
1) Perform recalibration of the “CAL BATT CHARGER” portion. 2) Replace the Battery Charger Board.
Parts Areas to Check
Battery Charger Board; Battery
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 12
Error Code System Message
ERROR 12 - Left drive stall / release-handle fault
Question
What does ERROR 12 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 12: Left drive stall / release-handle fault.
Possible Causes
Left stall signal detected.
Troubleshooting Steps
Check motor relays/snubber circuits, brakes, drive motor and stall/release wiring.
Parts Areas to Check
Drive control board; drive handle/Hall sensors; motor relays; drive motors; feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 12
Error Code System Message
12
Question
What should I check for error 13 on the AMX-4 Series?
Short Answer
Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards
Possible Causes
Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards
Troubleshooting Steps
Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards
Parts Areas to Check
Drive Amplifier Board
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 13
Error Code System Message
13
Question
What should I check for error 13 on the AMX-4 Series?
Short Answer
Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards
Possible Causes
Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards
Troubleshooting Steps
Note: This error displays as a left- or right-drive stall. 1) Check the drive motors for proper operation. 2) Make sure the brakes are fully released. 3) Check the Drive Control and Drive Amplifier Boards
Parts Areas to Check
Drive Amplifier Board
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 13
Error Code System Message
ERROR 13 - Right drive stall / release-handle fault
Question
What does ERROR 13 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 13: Right drive stall / release-handle fault.
Possible Causes
Right stall signal detected.
Troubleshooting Steps
Check right-side relay/snubber, brakes, drive motor and stall wiring.
Parts Areas to Check
Drive control board; drive handle/Hall sensors; motor relays; drive motors; feedback wiring