AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 8301053
Error Code System Message
Question
What does error 8301053 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301053: System battery failure – battery pack 2 temperature communication lost.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: ERROR.
Troubleshooting Steps
Check pack 2 temperature-sensor communication and battery connections; allow a cold system to stabilize before parts replacement.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301054
Error Code System Message
Question
What does error 8301054 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301054: System battery failure – battery pack 3 temperature communication lost.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: ERROR.
Troubleshooting Steps
Check pack 3 temperature-sensor communication and battery connections; allow a cold system to stabilize before parts replacement.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301055
Error Code System Message
Question
What does error 8301055 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301055: System battery failure – battery pack 4 temperature communication lost.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: ERROR.
Troubleshooting Steps
Check pack 4 temperature-sensor communication and battery connections; allow a cold system to stabilize before parts replacement.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301058
Error Code System Message
Question
What does error 8301058 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301058: Battery pack 1 temperature over 55C – charge inhibited.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Stop charging and allow cooling; evaluate pack condition and charging behavior.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301059
Error Code System Message
Question
What does error 8301059 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301059: Battery pack 2 temperature over 55C – charge inhibited.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Stop charging and allow cooling; evaluate pack condition and charging behavior.
Parts Areas to Check