AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 8301060
Error Code System Message
Question
What does error 8301060 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301060: Battery pack 3 temperature over 55C – charge inhibited.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Stop charging and allow cooling; evaluate pack condition and charging behavior.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301061
Error Code System Message
Question
What does error 8301061 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301061: Battery pack 4 temperature over 55C – charge inhibited.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Stop charging and allow cooling; evaluate pack condition and charging behavior.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301062
Error Code System Message
Question
What does error 8301062 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301062: Battery pack 1 temperature over 60C – breaker will trip.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: TRIP.
Troubleshooting Steps
Stop operation, allow cooling and evaluate pack 1 condition/temperature behavior.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301063
Error Code System Message
Question
What does error 8301063 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301063: Battery pack 2 temperature over 60C – breaker will trip.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: TRIP.
Troubleshooting Steps
Stop operation, allow cooling and evaluate pack 2 condition/temperature behavior.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
Error Number
Error 8301064
Error Code System Message
Question
What does error 8301064 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301064: Battery pack 3 temperature over 60C – breaker will trip.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: TRIP.
Troubleshooting Steps
Stop operation, allow cooling and evaluate pack 3 condition/temperature behavior.
Parts Areas to Check