top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 8301075

Error Code System Message

​

8301075, error 8301075, XR200 8301075, XR220 8301075, Optima XR200amx error 8301075, Optima XR220amx error 8301075, Battery stack voltage over 190V – breaker to trip
Question

What does error 8301075 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301075: Battery stack voltage over 190V – breaker to trip.

Possible Causes

Subsystem/category: System Battery. Manual recovery/error class: TRIP.

Troubleshooting Steps

Evaluate charging regulation, pack mismatch and connections.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Error 8301078

Error Code System Message

​

8301078, error 8301078, XR200 8301078, XR220 8301078, Optima XR200amx error 8301078, Optima XR220amx error 8301078, Average battery temperature over 55C
Question

What does error 8301078 mean on a GE Optima XR200amx or XR220amx?

Short Answer

Error 8301078: Average battery temperature over 55C.

Possible Causes

Subsystem/category: System Battery. Manual recovery/error class: WARNING.

Troubleshooting Steps

Allow cooling and evaluate battery-pack temperatures and charging conditions.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Errors 8301043, 8304044, 8304045, 8301046

Error Code System Message

​

8301043, 8301046, XR220 low battery breaker trip
Question

What do errors 8301043 and 8301046 mean on XR200amx or XR220amx?

Short Answer

They indicate low battery voltage: 8301043 commands breaker protection and 8301046 records low stack voltage below the idle/driving thresholds.

Possible Causes

Low state of charge; aged batteries with voltage sag under load; battery imbalance.

Troubleshooting Steps

Fully recharge the system before replacing parts. Review individual pack voltages and Power Control Limits. Repeated recurrence after a proper full charge suggests weak batteries.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Errors 8301058, 8301059, 8301060, 8301061

Error Code System Message

​

8301058, 8301059, 8301078, battery temperature charge inhibited
Question

What do XR200amx/XR220amx errors 8301058 through 8301061 mean?

Short Answer

These errors indicate an individual battery-pack temperature over 55°C during charging; 8301078 reports average battery temperature over 55°C.

Possible Causes

Weak high-impedance battery; high ambient temperature; battery/charging imbalance.

Troubleshooting Steps

Stop charging until temperatures are within limits. Inspect pack condition and charging behavior. Recurring overheating can indicate a weak pack.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Errors 8301062, 8301063, 8301064, 8301065

Error Code System Message

​

8301062, 8301065, 8301048, battery temp breaker trip
Question

What do errors 8301062 through 8301065 mean on XR200amx/XR220amx?

Short Answer

They indicate battery-pack temperature over 60°C and a breaker-trip condition.

Possible Causes

Overheated or weak battery pack; thermal transfer after charging; high ambient temperature.

Troubleshooting Steps

Allow the system/batteries to cool, review recent charging behavior and compare pack temperatures. Investigate a repeatedly hot pack before returning the system to service.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page