top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 19

Error Code System Message

ERROR 19 - Invalid drive command feedback

ERROR 19, Invalid drive command feedback, AMX-4 ERROR 19, AMX-4+ ERROR 19, Definium AMX 700 ERROR 19, GE AMX error ERROR 19
Question

What does ERROR 19 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 19: Invalid drive command feedback.

Possible Causes

Drive-command feedback was invalid.

Troubleshooting Steps

Check drive fuses, connectors and drive-control circuitry; investigate intermittent command feedback.

Parts Areas to Check

Drive control board; drive handle/Hall sensors; motor relays; drive motors; feedback wiring

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 19

Error Code System Message

19

AMX-4, AMX-4+, Definium AMX 700, ERROR 19, Error 19
Question

What should I check for error 19 on the AMX-4 Series?

Short Answer

1) Usually caused when the unit travels from carpet to tile, enters an elevator, or is driven over uneven tile flooring. This is generally a non-fatal error.

Possible Causes

Usually caused when the unit travels from carpet to tile, enters an elevator, or is driven over uneven tile flooring. This is generally a non-fatal error.

Troubleshooting Steps

1) Usually caused when the unit travels from carpet to tile, enters an elevator, or is driven over uneven tile flooring. This is generally a non-fatal error.

Parts Areas to Check

​

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 210

Error Code System Message

210

AMX-4, AMX-4+, Definium AMX 700, ERROR 210, Error 210
Question

What should I check for errors 210 on the AMX-4 Series?

Short Answer

1) Check the Drive Controller Board for proper feedback

Possible Causes

Check the Drive Controller Board for proper feedback

Troubleshooting Steps

1) Check the Drive Controller Board for proper feedback

Parts Areas to Check

​

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 210

Error Code System Message

ERROR 210 - Left stall feedback circuit fault

ERROR 210, Left stall feedback circuit fault, AMX-4 ERROR 210, AMX-4+ ERROR 210, Definium AMX 700 ERROR 210, GE AMX error ERROR 210
Question

What does ERROR 210 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 210: Left stall feedback circuit fault.

Possible Causes

Left stall-feedback circuitry appears defective.

Troubleshooting Steps

Check drive-board-to-CPU connection and left stall feedback.

Parts Areas to Check

Drive control board; drive handle/Hall sensors; motor relays; drive motors; feedback wiring

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 211

Error Code System Message

ERROR 211 - Right stall feedback circuit fault

ERROR 211, Right stall feedback circuit fault, AMX-4 ERROR 211, AMX-4+ ERROR 211, Definium AMX 700 ERROR 211, GE AMX error ERROR 211
Question

What does ERROR 211 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 211: Right stall feedback circuit fault.

Possible Causes

Right stall-feedback circuitry appears defective.

Troubleshooting Steps

Check drive-board-to-CPU connection and right stall feedback.

Parts Areas to Check

Drive control board; drive handle/Hall sensors; motor relays; drive motors; feedback wiring

For Related Models

AMX-4, AMX-4+, Definium AMX 700