top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Hospital network communication issues

Error Code System Message

​

XR200 PACS network, XR220 hospital network, TCP IP diagnostic
Question

Which diagnostic checks hospital-network communication on XR200amx and XR220amx?

Short Answer

The Hospital Network diagnostic performs low-level TCP/IP checks; Port to Port Diag isolates PC ports to test transmit/receive functionality.

Possible Causes

TCP/IP configuration; hospital network access; PC port/cabling problem.

Troubleshooting Steps

Run Hospital Network first. If needed, use Port to Port Diag to isolate PC port hardware, and coordinate with hospital IT for network addressing or access issues.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

LVLE2 power issues

Error Code System Message

​

XR200 LVLE2 problem, XR220 power distribution, LVLE2 current
Question

How do I isolate an LVLE2 power problem on XR200amx or XR220amx?

Short Answer

The Exclusive manual recommends testing LVLE2 under load and distinguishing a supply fault from an excessive-current load or cabling problem.

Possible Causes

LVLE2 failure; downstream board/subsystem fault; cabling short or excessive load.

Troubleshooting Steps

Reconnect loads, run Boundary Scan where available, and evaluate LVLE2 output under load. If a specific output shows excessive current, isolate the corresponding load/cabling rather than replacing LVLE2 immediately.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

No primary dector selected message

Error Code System Message

​

XR220 no primary detector, active detector, Detector Management
Question

What does “No Primary Detector selected” mean on XR220amx?

Short Answer

The detector icon indicates that the user has not selected an active/primary detector for imaging.

Possible Causes

No detector selected as Active in Detector Management.

Troubleshooting Steps

Open Detector Management and select the desired registered detector as Active. Only one detector can be Active at a time.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Power control limits

Error Code System Message

​

XR220 Power Control Limits healthy, battery pack voltages
Question

What does a healthy XR200amx/XR220amx Power Control Limits test look like?

Short Answer

A healthy set should show balanced individual pack voltages, reasonable idle battery/LVLE2 current and normal pack temperatures, with the diagnostic passing.

Possible Causes

Diagnostic interpretation question.

Troubleshooting Steps

Compare all four pack voltages rather than only total stack voltage. Look for one pack that differs significantly, abnormal current, or elevated temperature.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

Error Number

Power problem

Error Code System Message

​

XR200 power problem, XR220 power diagnostic, Power Control Limits
Question

Which diagnostics should I run for an XR200amx or XR220amx power problem?

Short Answer

Use the System Power diagnostic group to check power-control communication, battery voltage/current limits, thermal readings, breaker control and AC-present status.

Possible Causes

Power-control communication fault; low or abnormal battery voltage/current; thermal condition; breaker or AC detection issue.

Troubleshooting Steps

Run Power Control Comm and Power Control Limits first. Use Power Supply Thermal, Breaker Trip and AC Present when the symptom points to those functions.

Parts Areas to Check

​

For Related Models

Optima XR200amx, Optima XR220amx

bottom of page