Error Number

Detector won't pair in the charging bin

Error Code System Message

​

XR240 detector pair failed, detector not pairing, FlashPad pairing, detector charging bin
Question

Why won't my GE Optima XR240amx detector pair in the charging bin?

Short Answer

The detector may be unregistered, incorrectly configured, or unable to communicate through the wireless access point.

Possible Causes

Detector not registered; AP configuration issue; unsupported wireless country/channel combination.

Troubleshooting Steps

Check the registered-detector list. Register the detector if needed, then insert it in the proper bin to pair. If AP configuration messages appear, verify the selected country configuration and AP operation.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0100

Error Code System Message

​

0100, error 0100, XR240 0100, Optima XR240amx error 0100, Unexpected generator firmware error
Question

What does error 0100 mean on a GE Optima XR240amx?

Short Answer

Error 0100: Unexpected generator firmware error.

Possible Causes

Subsystem/category: Fil Small. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0101

Error Code System Message

​

0101, error 0101, XR240 0101, Optima XR240amx error 0101, Filament overcurrent detected
Question

What does error 0101 mean on a GE Optima XR240amx?

Short Answer

Error 0101: Filament overcurrent detected.

Possible Causes

Subsystem/category: Fil Small. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0102

Error Code System Message

​

0102, error 0102, XR240 0102, Optima XR240amx error 0102, Filament open circuit detected
Question

What does error 0102 mean on a GE Optima XR240amx?

Short Answer

Error 0102: Filament open circuit detected.

Possible Causes

Subsystem/category: Fil Small. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0103

Error Code System Message

​

0103, error 0103, XR240 0103, Optima XR240amx error 0103, Filament drive current exceeds maximum limit
Question

What does error 0103 mean on a GE Optima XR240amx?

Short Answer

Error 0103: Filament drive current exceeds maximum limit.

Possible Causes

Subsystem/category: Fil Small. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR240amx

bottom of page