AMX Troubleshooting & Technical Support
Results Count
Error Number
Detector won't pair in the charging bin
Error Code System Message
Question
Why won't my GE Optima XR240amx detector pair in the charging bin?
Short Answer
The detector may be unregistered, incorrectly configured, or unable to communicate through the wireless access point.
Possible Causes
Detector not registered; AP configuration issue; unsupported wireless country/channel combination.
Troubleshooting Steps
Check the registered-detector list. Register the detector if needed, then insert it in the proper bin to pair. If AP configuration messages appear, verify the selected country configuration and AP operation.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0100
Error Code System Message
Question
What does error 0100 mean on a GE Optima XR240amx?
Short Answer
Error 0100: Unexpected generator firmware error.
Possible Causes
Subsystem/category: Fil Small. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0101
Error Code System Message
Question
What does error 0101 mean on a GE Optima XR240amx?
Short Answer
Error 0101: Filament overcurrent detected.
Possible Causes
Subsystem/category: Fil Small. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0102
Error Code System Message
Question
What does error 0102 mean on a GE Optima XR240amx?
Short Answer
Error 0102: Filament open circuit detected.
Possible Causes
Subsystem/category: Fil Small. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0103
Error Code System Message
Question
What does error 0103 mean on a GE Optima XR240amx?
Short Answer
Error 0103: Filament drive current exceeds maximum limit.
Possible Causes
Subsystem/category: Fil Small. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check