AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0183
Error Code System Message
Question
What does error 0183 mean on a GE Optima XR240amx?
Short Answer
Error 0183: Auto diagnostic failed during generator boot.
Possible Causes
Subsystem/category: Fil Small. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0200
Error Code System Message
Question
What does error 0200 mean on a GE Optima XR240amx?
Short Answer
Error 0200: Unexpected generator firmware error.
Possible Causes
Subsystem/category: Fil Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0201
Error Code System Message
Question
What does error 0201 mean on a GE Optima XR240amx?
Short Answer
Error 0201: Filament overcurrent detected.
Possible Causes
Subsystem/category: Fil Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0202
Error Code System Message
Question
What does error 0202 mean on a GE Optima XR240amx?
Short Answer
Error 0202: Filament open circuit detected.
Possible Causes
Subsystem/category: Fil Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0203
Error Code System Message
Question
What does error 0203 mean on a GE Optima XR240amx?
Short Answer
Error 0203: Filament drive current exceeds maximum limit.
Possible Causes
Subsystem/category: Fil Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check