AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0412
Error Code System Message
Question
What does error 0412 mean on a GE Optima XR240amx?
Short Answer
Error 0412: Generator software corruption at boot.
Possible Causes
Subsystem/category: Rotor. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics. Inspect rotor communication, tube stator/harness connections, and Power/Main Auxiliary circuitry.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0413
Error Code System Message
Question
What does error 0413 mean on a GE Optima XR240amx?
Short Answer
Error 0413: Overcurrent detected on rotation phases.
Possible Causes
Subsystem/category: Rotor. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics. Inspect rotor communication, tube stator/harness connections, and Power/Main Auxiliary circuitry.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0458
Error Code System Message
Question
What does error 0458 mean on a GE Optima XR240amx?
Short Answer
Error 0458: Communication failure between Power and Main Auxiliary modules.
Possible Causes
Subsystem/category: Rotor. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics. Inspect rotor communication, tube stator/harness connections, and Power/Main Auxiliary circuitry.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0460
Error Code System Message
Question
What does error 0460 mean on a GE Optima XR240amx?
Short Answer
Error 0460: Communication failure between Power and Main Auxiliary modules.
Possible Causes
Subsystem/category: Rotor. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics. Inspect rotor communication, tube stator/harness connections, and Power/Main Auxiliary circuitry.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0462
Error Code System Message
Question
What does error 0462 mean on a GE Optima XR240amx?
Short Answer
Error 0462: Communication failure between Power and Main Auxiliary modules.
Possible Causes
Subsystem/category: Rotor. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics. Inspect rotor communication, tube stator/harness connections, and Power/Main Auxiliary circuitry.
Parts Areas to Check