Error Number
Error 0504
Error Code System Message
Question
What does error 0504 mean on a GE Optima XR240amx?
Short Answer
Error 0504: DC-bus rise time too slow during DC-bus charge.
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0505
Error Code System Message
Question
What does error 0505 mean on a GE Optima XR240amx?
Short Answer
Error 0505: No input current detected during DC-bus charge.
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0510
Error Code System Message
Question
What does error 0510 mean on a GE Optima XR240amx?
Short Answer
Error 0510: Overvoltage on DC output detected (DC bus >460 V).
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0511
Error Code System Message
Question
What does error 0511 mean on a GE Optima XR240amx?
Short Answer
Error 0511: Unexpected generator firmware error.
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0520
Error Code System Message
Question
What does error 0520 mean on a GE Optima XR240amx?
Short Answer
Error 0520: Software over-input current detected (Iin >140 A).
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.
Parts Areas to Check