top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0504

Error Code System Message

​

0504, error 0504, XR240 0504, Optima XR240amx error 0504, DC-bus rise time too slow during DC-bus charge
Question

What does error 0504 mean on a GE Optima XR240amx?

Short Answer

Error 0504: DC-bus rise time too slow during DC-bus charge.

Possible Causes

Subsystem/category: Input Stage. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0505

Error Code System Message

​

0505, error 0505, XR240 0505, Optima XR240amx error 0505, No input current detected during DC-bus charge
Question

What does error 0505 mean on a GE Optima XR240amx?

Short Answer

Error 0505: No input current detected during DC-bus charge.

Possible Causes

Subsystem/category: Input Stage. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0510

Error Code System Message

​

0510, error 0510, XR240 0510, Optima XR240amx error 0510, Overvoltage on DC output detected (DC bus >460 V)
Question

What does error 0510 mean on a GE Optima XR240amx?

Short Answer

Error 0510: Overvoltage on DC output detected (DC bus >460 V).

Possible Causes

Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0511

Error Code System Message

​

0511, error 0511, XR240 0511, Optima XR240amx error 0511, Unexpected generator firmware error
Question

What does error 0511 mean on a GE Optima XR240amx?

Short Answer

Error 0511: Unexpected generator firmware error.

Possible Causes

Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0520

Error Code System Message

​

0520, error 0520, XR240 0520, Optima XR240amx error 0520, Software over-input current detected (Iin >140 A)
Question

What does error 0520 mean on a GE Optima XR240amx?

Short Answer

Error 0520: Software over-input current detected (Iin >140 A).

Possible Causes

Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.

Parts Areas to Check

​

For Related Models

Optima XR240amx

bottom of page