AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0581

Error Code System Message

​

XR240 error 0581, low input voltage generator, battery boost error
Question

What does XR240amx generator error 0581 mean?

Short Answer

Error 0581 indicates low input voltage during generator boost and can point to system battery condition or the generator input-power path.

Possible Causes

Low battery state of charge; input power connection problem; battery stack/cabling issue.

Troubleshooting Steps

Check system battery condition and Power Control Limits values. Inspect the generator input power path and battery-stack connection as specified by the service manual. Fully charge low batteries before concluding a generator component is faulty.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0582

Error Code System Message

​

0582, error 0582, XR240 0582, Optima XR240amx error 0582, High input voltage during boost (Vin >200 V)
Question

What does error 0582 mean on a GE Optima XR240amx?

Short Answer

Error 0582: High input voltage during boost (Vin >200 V).

Possible Causes

Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0582

Error Code System Message

​

XR240 error 0582, high input voltage boost, Djinn generator
Question

What does XR240amx generator error 0582 mean?

Short Answer

Error 0582 is listed as high input voltage during boost.

Possible Causes

Abnormal generator input voltage; battery/input measurement issue.

Troubleshooting Steps

Use the manual's input-voltage measurement and generator power-path checks to verify the condition before replacing parts.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0583

Error Code System Message

​

0583, error 0583, XR240 0583, Optima XR240amx error 0583, Low DC-bus voltage during boost (Vout
Question

What does error 0583 mean on a GE Optima XR240amx?

Short Answer

Error 0583: Low DC-bus voltage during boost (Vout <280 V).

Possible Causes

Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0584

Error Code System Message

​

0584, error 0584, XR240 0584, Optima XR240amx error 0584, No input voltage detected before boost start
Question

What does error 0584 mean on a GE Optima XR240amx?

Short Answer

Error 0584: No input voltage detected before boost start.

Possible Causes

Subsystem/category: Input Stage. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.

Parts Areas to Check

​

For Related Models

Optima XR240amx