AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0585
Error Code System Message
Question
What does error 0585 mean on a GE Optima XR240amx?
Short Answer
Error 0585: Input current detected while boost is off (Iin >8 A).
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0600
Error Code System Message
Question
What does error 0600 mean on a GE Optima XR240amx?
Short Answer
Error 0600: Unexpected generator firmware error.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0601
Error Code System Message
Question
What does error 0601 mean on a GE Optima XR240amx?
Short Answer
Error 0601: Exposure sequence error from system.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0602
Error Code System Message
Question
What does error 0602 mean on a GE Optima XR240amx?
Short Answer
Error 0602: Unbalanced kV: cathode-to-anode difference exceeds 10 kV.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0603
Error Code System Message
Question
What does error 0603 mean on a GE Optima XR240amx?
Short Answer
Error 0603: Unbalanced kV: anode-to-cathode difference exceeds 10 kV.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check