Error 0585

Error Code System Message

​

0585, error 0585, XR240 0585, Optima XR240amx error 0585, Input current detected while boost is off (Iin >8 A)
Question

What does error 0585 mean on a GE Optima XR240amx?

Short Answer

Error 0585: Input current detected while boost is off (Iin >8 A).

Possible Causes

Subsystem/category: Input Stage. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0600

Error Code System Message

​

0600, error 0600, XR240 0600, Optima XR240amx error 0600, Unexpected generator firmware error
Question

What does error 0600 mean on a GE Optima XR240amx?

Short Answer

Error 0600: Unexpected generator firmware error.

Possible Causes

Subsystem/category: kV. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0601

Error Code System Message

​

0601, error 0601, XR240 0601, Optima XR240amx error 0601, Exposure sequence error from system
Question

What does error 0601 mean on a GE Optima XR240amx?

Short Answer

Error 0601: Exposure sequence error from system.

Possible Causes

Subsystem/category: kV. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0602

Error Code System Message

​

0602, error 0602, XR240 0602, Optima XR240amx error 0602, Unbalanced kV: cathode-to-anode difference exceeds 10 kV
Question

What does error 0602 mean on a GE Optima XR240amx?

Short Answer

Error 0602: Unbalanced kV: cathode-to-anode difference exceeds 10 kV.

Possible Causes

Subsystem/category: kV. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0603

Error Code System Message

​

0603, error 0603, XR240 0603, Optima XR240amx error 0603, Unbalanced kV: anode-to-cathode difference exceeds 10 kV
Question

What does error 0603 mean on a GE Optima XR240amx?

Short Answer

Error 0603: Unbalanced kV: anode-to-cathode difference exceeds 10 kV.

Possible Causes

Subsystem/category: kV. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​