top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 060A

Error Code System Message

​

060A, error 060A, XR240 060A, Optima XR240amx error 060A, Measured kV too low after exposure start (kV
Question

What does error 060A mean on a GE Optima XR240amx?

Short Answer

Error 060A: Measured kV too low after exposure start (kV <3 kV).

Possible Causes

Subsystem/category: kV. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 060C

Error Code System Message

​

060C, error 060C, XR240 060C, Optima XR240amx error 060C, Unexpected generator firmware error
Question

What does error 060C mean on a GE Optima XR240amx?

Short Answer

Error 060C: Unexpected generator firmware error.

Possible Causes

Subsystem/category: kV. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 060E

Error Code System Message

​

060E, error 060E, XR240 060E, Optima XR240amx error 060E, Measured kV too low during exposure (
Question

What does error 060E mean on a GE Optima XR240amx?

Short Answer

Error 060E: Measured kV too low during exposure (<75% of kV target).

Possible Causes

Subsystem/category: kV. Manual recovery/error class: ABORT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 060F

Error Code System Message

​

060F, error 060F, XR240 060F, Optima XR240amx error 060F, Inverter current regulation issue
Question

What does error 060F mean on a GE Optima XR240amx?

Short Answer

Error 060F: Inverter current regulation issue.

Possible Causes

Subsystem/category: kV. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0611

Error Code System Message

​

0611, error 0611, XR240 0611, Optima XR240amx error 0611, Issue detected on X-ray hardware command (RTL2)
Question

What does error 0611 mean on a GE Optima XR240amx?

Short Answer

Error 0611: Issue detected on X-ray hardware command (RTL2).

Possible Causes

Subsystem/category: RTL. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review the Class M / generator error log for related errors first. Restart only when appropriate, then verify generator software integrity, communications, power and cabling using the applicable SUIF diagnostics.

Parts Areas to Check

​

For Related Models

Optima XR240amx

bottom of page