AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 060A
Error Code System Message
Question
What does error 060A mean on a GE Optima XR240amx?
Short Answer
Error 060A: Measured kV too low after exposure start (kV <3 kV).
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 060C
Error Code System Message
Question
What does error 060C mean on a GE Optima XR240amx?
Short Answer
Error 060C: Unexpected generator firmware error.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 060E
Error Code System Message
Question
What does error 060E mean on a GE Optima XR240amx?
Short Answer
Error 060E: Measured kV too low during exposure (<75% of kV target).
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 060F
Error Code System Message
Question
What does error 060F mean on a GE Optima XR240amx?
Short Answer
Error 060F: Inverter current regulation issue.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0611
Error Code System Message
Question
What does error 0611 mean on a GE Optima XR240amx?
Short Answer
Error 0611: Issue detected on X-ray hardware command (RTL2).
Possible Causes
Subsystem/category: RTL. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the Class M / generator error log for related errors first. Restart only when appropriate, then verify generator software integrity, communications, power and cabling using the applicable SUIF diagnostics.
Parts Areas to Check