AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0613
Error Code System Message
Question
What does error 0613 mean on a GE Optima XR240amx?
Short Answer
Error 0613: Issue detected on X-ray hardware enable (RTL4).
Possible Causes
Subsystem/category: RTL. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the Class M / generator error log for related errors first. Restart only when appropriate, then verify generator software integrity, communications, power and cabling using the applicable SUIF diagnostics.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0616
Error Code System Message
Question
What does error 0616 mean on a GE Optima XR240amx?
Short Answer
Error 0616: Demanded techniques cannot be achieved.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0681
Error Code System Message
Question
What does error 0681 mean on a GE Optima XR240amx?
Short Answer
Error 0681: Synchronization issue at exposure start.
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0684
Error Code System Message
Question
What does error 0684 mean on a GE Optima XR240amx?
Short Answer
Error 0684: kV detected while inverter is off (>1 kV).
Possible Causes
Subsystem/category: kV. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0685
Error Code System Message
Question
What does error 0685 mean on a GE Optima XR240amx?
Short Answer
Error 0685: Resonant current detected while inverter is off (ILR >1 A).
Possible Causes
Subsystem/category: kV. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check