AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0701
Error Code System Message
Question
What does error 0701 mean on a GE Optima XR240amx?
Short Answer
Error 0701: Tube over-temperature switch open.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0702
Error Code System Message
Question
What does error 0702 mean on a GE Optima XR240amx?
Short Answer
Error 0702: Tank temperature is too high.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0703
Error Code System Message
Question
What does error 0703 mean on a GE Optima XR240amx?
Short Answer
Error 0703: Power board temperature is too high.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0704
Error Code System Message
Question
What does error 0704 mean on a GE Optima XR240amx?
Short Answer
Error 0704: Filament board temperature is too high.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0705
Error Code System Message
Question
What does error 0705 mean on a GE Optima XR240amx?
Short Answer
Error 0705: Rotor board temperature is too high.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.
Parts Areas to Check