AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0706
Error Code System Message
Question
What does error 0706 mean on a GE Optima XR240amx?
Short Answer
Error 0706: Tank temperature is getting hot.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0707
Error Code System Message
Question
What does error 0707 mean on a GE Optima XR240amx?
Short Answer
Error 0707: Power board temperature is rising.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0708
Error Code System Message
Question
What does error 0708 mean on a GE Optima XR240amx?
Short Answer
Error 0708: Filament board temperature is rising.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0709
Error Code System Message
Question
What does error 0709 mean on a GE Optima XR240amx?
Short Answer
Error 0709: Rotor board temperature is rising.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0782
Error Code System Message
Question
What does error 0782 mean on a GE Optima XR240amx?
Short Answer
Error 0782: Tank temperature sensor failure detected.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.
Parts Areas to Check