top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0783

Error Code System Message

​

0783, error 0783, XR240 0783, Optima XR240amx error 0783, Power board temperature sensor failure
Question

What does error 0783 mean on a GE Optima XR240amx?

Short Answer

Error 0783: Power board temperature sensor failure.

Possible Causes

Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0784

Error Code System Message

​

0784, error 0784, XR240 0784, Optima XR240amx error 0784, Tube thermal protection error
Question

What does error 0784 mean on a GE Optima XR240amx?

Short Answer

Error 0784: Tube thermal protection error.

Possible Causes

Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Stop repeated exposures if overheating is suspected. Allow cooling, verify airflow, and run Power Supply Thermal / generator thermal diagnostics. Inspect fans, sensors and the identified generator board.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0785

Error Code System Message

​

0785, error 0785, XR240 0785, Optima XR240amx error 0785, Revision/device configuration failure detected
Question

What does error 0785 mean on a GE Optima XR240amx?

Short Answer

Error 0785: Revision/device configuration failure detected.

Possible Causes

Subsystem/category: Config. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review generator configuration/software state and related errors. Verify generator parts/configuration and software compatibility before replacing hardware.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0786

Error Code System Message

​

0786, error 0786, XR240 0786, Optima XR240amx error 0786, Unexpected generator parts detected
Question

What does error 0786 mean on a GE Optima XR240amx?

Short Answer

Error 0786: Unexpected generator parts detected.

Possible Causes

Subsystem/category: Config. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review generator configuration/software state and related errors. Verify generator parts/configuration and software compatibility before replacing hardware.

Parts Areas to Check

​

For Related Models

Optima XR240amx

Error Number

Error 0787

Error Code System Message

​

0787, error 0787, XR240 0787, Optima XR240amx error 0787, Generator parts identification is impossible
Question

What does error 0787 mean on a GE Optima XR240amx?

Short Answer

Error 0787: Generator parts identification is impossible.

Possible Causes

Subsystem/category: Config. Manual recovery/error class: INHIBIT.

Troubleshooting Steps

Review generator configuration/software state and related errors. Verify generator parts/configuration and software compatibility before replacing hardware.

Parts Areas to Check

​

For Related Models

Optima XR240amx

bottom of page