AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0A85
Error Code System Message
Question
What does error 0A85 mean on a GE Optima XR240amx?
Short Answer
Error 0A85: External 48 V power supplies out of specification.
Possible Causes
Subsystem/category: Misc. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the Class M / generator error log for related errors first. Restart only when appropriate, then verify generator software integrity, communications, power and cabling using the applicable SUIF diagnostics.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0A87
Error Code System Message
Question
What does error 0A87 mean on a GE Optima XR240amx?
Short Answer
Error 0A87: Simulation software detected.
Possible Causes
Subsystem/category: Misc. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the Class M / generator error log for related errors first. Restart only when appropriate, then verify generator software integrity, communications, power and cabling using the applicable SUIF diagnostics.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0AA0
Error Code System Message
Question
What does error 0AA0 mean on a GE Optima XR240amx?
Short Answer
Error 0AA0: Reboot caused by an unexpected crash.
Possible Causes
Subsystem/category: Misc. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review the Class M / generator error log for related errors first. Restart only when appropriate, then verify generator software integrity, communications, power and cabling using the applicable SUIF diagnostics.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0AA1
Error Code System Message
Question
What does error 0AA1 mean on a GE Optima XR240amx?
Short Answer
Error 0AA1: Watchdog software error.
Possible Causes
Subsystem/category: WDT. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the Class M / generator error log for related errors first. Restart only when appropriate, then verify generator software integrity, communications, power and cabling using the applicable SUIF diagnostics.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 0B81
Error Code System Message
Question
What does error 0B81 mean on a GE Optima XR240amx?
Short Answer
Error 0B81: mA measured while inverter is off (>1 mA).
Possible Causes
Subsystem/category: mA. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check