AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0B87
Error Code System Message
Question
What does error 0B87 mean on a GE Optima XR240amx?
Short Answer
Error 0B87: Measured mA is out of range during exposure (<10% of target).
Possible Causes
Subsystem/category: mA. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 122102
Error Code System Message
Question
What does error 122102 mean on a GE Optima XR240amx?
Short Answer
Error 122102: Digital Cassette Detector Boot is in progress.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
Allow detector boot to complete. If it never reaches boot complete/calibration upload complete, troubleshoot wireless or tether communication.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221030
Error Code System Message
Question
What does error 1221030 mean on a GE Optima XR240amx?
Short Answer
Error 1221030: Digital Cassette Detector Boot is complete.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
This is a successful status message; continue to calibration upload/registration as applicable.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221036
Error Code System Message
Question
What does error 1221036 mean on a GE Optima XR240amx?
Short Answer
Error 1221036: Digital Cassette firmware is incorrect.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
Connect by tether if available, verify firmware with Built In Self Test, and load the correct firmware/compatibility table using the approved service procedure.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221101
Error Code System Message
Question
What does error 1221101 mean on a GE Optima XR240amx?
Short Answer
Error 1221101: Detector calibration upload in progress.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
Wait for calibration upload to complete before registration or imaging workflow continues.
Parts Areas to Check