AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 1221233
Error Code System Message
Question
What does error 1221233 mean on a GE Optima XR240amx?
Short Answer
Error 1221233: Successful pairing of the detector.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: ADVISORY.
Troubleshooting Steps
No corrective action is required; allow boot and calibration upload sequence to finish.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221234
Error Code System Message
Question
What does error 1221234 mean on a GE Optima XR240amx?
Short Answer
Error 1221234: Pairing of the detector failed.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
Confirm detector registration, bin placement, AP communication and country/channel configuration, then retry pairing.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221235
Error Code System Message
Question
What does error 1221235 mean on a GE Optima XR240amx?
Short Answer
Error 1221235: Successful pairing of detector via tether.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
No corrective action is required; allow boot/calibration sequence to finish.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221239
Error Code System Message
Question
What does error 1221239 mean on a GE Optima XR240amx?
Short Answer
Error 1221239: Successful pairing of detector in large bin slot.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
No corrective action is required; allow the successful boot and calibration upload sequence to complete.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 1221240
Error Code System Message
Question
What does error 1221240 mean on a GE Optima XR240amx?
Short Answer
Error 1221240: Successful pairing of detector in small bin slot.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
No corrective action is required; allow the successful boot and calibration upload sequence to complete.
Parts Areas to Check