AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 8110
Error Code System Message
Question
What does error 8110 mean on a GE Optima XR240amx?
Short Answer
Error 8110: Communication overflow detected.
Possible Causes
Subsystem/category: CANopen. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review the Class M / generator error log for related errors first. Restart only when appropriate, then verify generator software integrity, communications, power and cabling using the applicable SUIF diagnostics.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Error 8130
Error Code System Message
Question
What does error 8130 mean on a GE Optima XR240amx?
Short Answer
Error 8130: Communication watchdog failure detected.
Possible Causes
Subsystem/category: CANopen. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the Class M / generator error log for related errors first. Restart only when appropriate, then verify generator software integrity, communications, power and cabling using the applicable SUIF diagnostics.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
First troubleshooting step
Error Code System Message
Question
What is the first troubleshooting step for a GE Optima XR240amx?
Short Answer
The manual says to review the error log first, speak with the operator about the circumstances, then test the affected system area using the applicable diagnostics.
Possible Causes
Any system fault.
Troubleshooting Steps
1) Review the error log to identify affected functionality. 2) Ask the operator when/how the issue occurs. 3) Run the applicable SUIF diagnostic for the affected subsystem.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Generator Communication Failure
Error Code System Message
Question
What causes “Generator Communication Failure” on the XR240amx?
Short Answer
The manual treats this as a Djinn generator communication problem and recommends reviewing recent generator errors and related system issues before ordering parts.
Possible Causes
Lost generator communication; rotor communication problem; generator in error state; wiring/power issue.
Troubleshooting Steps
Review the Class M error log for recent generator errors. Check related system errors, generator power/communications and wiring. Restart the system/generator if appropriate, then run the applicable generator diagnostics.
Parts Areas to Check
For Related Models
Optima XR240amx
Error Number
Generator diagnostics
Error Code System Message
Question
Which generator diagnostics are available on the XR240amx?
Short Answer
The manual lists Rotation, Tube Rotor Impedance, Power Supplies, Resonant Components, Inverter, 70C Circuit, Rise Time/Regulation, filament-driver, mA, kV and Auto tests.
Possible Causes
Generator/exposure subsystem fault.
Troubleshooting Steps
Select the generator diagnostic that corresponds to the error log and symptom. Use compound kV, mA or Auto tests when appropriate and follow the service manual sequence for any failed subtest.
Parts Areas to Check