AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 447
Error Code System Message
ERROR 447 - 60 Hz inverter ON signal missing
Question
What does ERROR 447 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 447: 60 Hz inverter ON signal missing.
Possible Causes
The 60 Hz inverter ON signal did not assert when expected.
Troubleshooting Steps
Check inverter ON/relay signal path and generator control.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 448
Error Code System Message
ERROR 448 - Rotor interlock feedback missing after 60 Hz inverter start
Question
What does ERROR 448 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 448: Rotor interlock feedback missing after 60 Hz inverter start.
Possible Causes
Rotor interlock did not assert in the expected sequence.
Troubleshooting Steps
Check rotor interlock signal, relays and generator-controller timing.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 449
Error Code System Message
ERROR 449 - kVp command feedback low
Question
What does ERROR 449 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 449: kVp command feedback low.
Possible Causes
kVp command feedback was lower than commanded.
Troubleshooting Steps
Check filament/kVp feedback, CPU-generator connection and generator-control output.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 45
Error Code System Message
ERROR 45 - Battery voltage low during prep
Question
What does ERROR 45 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 45: Battery voltage low during prep.
Possible Causes
Battery voltage was lower than expected during prep.
Troubleshooting Steps
Recharge/evaluate battery set and verify battery-voltage calibration.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
Error 450
Error Code System Message
450
Question
What should I check for error 450 on the AMX-4 Series?
Short Answer
1. Check wiring to X-Ray Tube Assembly pressure switch (switch should be closed). If switch closure rings back to J3-6 & 7 of CPU Board, then problem on CPU Board
Possible Causes
1) Check wiring to X-Ray Tube Assembly pressure switch (switch should be closed). If switch closure rings back to J3-6 & 7 of CPU Board, then problem on CPU Board, 2) Bad X-Ray Tube
Troubleshooting Steps
1. Check wiring to X-Ray Tube Assembly pressure switch (switch should be closed). If switch closure rings back to J3-6 & 7 of CPU Board, then problem on CPU Board
Parts Areas to Check
CPU Board; X-Ray Tube