top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0104

Error Code System Message

​

0x0104, 0X0104, Brivo XR285amx 0x0104, Djinn Heater 1 error, Demand too low (target
Question

What does 0x0104 mean on a GE Brivo XR285amx?

Short Answer

0x0104: Demand too low (target < min heater).

Possible Causes

The Djinn Beetle generator reported a Heater 1 FW fault: Demand too low (target < min heater).

Troubleshooting Steps

Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0105

Error Code System Message

​

0x0105, 0X0105, Brivo XR285amx 0x0105, Djinn Heater 1 error, Inverter OFF and current not zero
Question

What does 0x0105 mean on a GE Brivo XR285amx?

Short Answer

0x0105: Inverter OFF and current not zero.

Possible Causes

The Djinn Beetle generator reported a Heater 1 FW fault: Inverter OFF and current not zero.

Troubleshooting Steps

Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0107

Error Code System Message

​

0x0107, 0X0107, Brivo XR285amx 0x0107, Djinn Heater 1 error, Over current detected by FPGA
Question

What does 0x0107 mean on a GE Brivo XR285amx?

Short Answer

0x0107: Over current detected by FPGA.

Possible Causes

The Djinn Beetle generator reported a Heater 1 FW fault: Over current detected by FPGA.

Troubleshooting Steps

Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0108

Error Code System Message

​

0x0108, 0X0108, Brivo XR285amx 0x0108, Djinn Heater 1 error, Filament 1 driver stopped by 48V out of range
Question

What does 0x0108 mean on a GE Brivo XR285amx?

Short Answer

0x0108: Filament 1 driver stopped by 48V out of range.

Possible Causes

The Djinn Beetle generator reported a Heater 1 FW fault: Filament 1 driver stopped by 48V out of range.

Troubleshooting Steps

Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0181

Error Code System Message

​

0x0181, 0X0181, Brivo XR285amx 0x0181, Djinn Heater 1 error, state SW Unexpected firmware state
Question

What does 0x0181 mean on a GE Brivo XR285amx?

Short Answer

0x0181: state SW Unexpected firmware state.

Possible Causes

The Djinn Beetle generator reported a Heater 1 SW fault: state SW Unexpected firmware state.

Troubleshooting Steps

Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page