AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0x0104
Error Code System Message
Question
What does 0x0104 mean on a GE Brivo XR285amx?
Short Answer
0x0104: Demand too low (target < min heater).
Possible Causes
The Djinn Beetle generator reported a Heater 1 FW fault: Demand too low (target < min heater).
Troubleshooting Steps
Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0105
Error Code System Message
Question
What does 0x0105 mean on a GE Brivo XR285amx?
Short Answer
0x0105: Inverter OFF and current not zero.
Possible Causes
The Djinn Beetle generator reported a Heater 1 FW fault: Inverter OFF and current not zero.
Troubleshooting Steps
Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0107
Error Code System Message
Question
What does 0x0107 mean on a GE Brivo XR285amx?
Short Answer
0x0107: Over current detected by FPGA.
Possible Causes
The Djinn Beetle generator reported a Heater 1 FW fault: Over current detected by FPGA.
Troubleshooting Steps
Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0108
Error Code System Message
Question
What does 0x0108 mean on a GE Brivo XR285amx?
Short Answer
0x0108: Filament 1 driver stopped by 48V out of range.
Possible Causes
The Djinn Beetle generator reported a Heater 1 FW fault: Filament 1 driver stopped by 48V out of range.
Troubleshooting Steps
Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0181
Error Code System Message
Question
What does 0x0181 mean on a GE Brivo XR285amx?
Short Answer
0x0181: state SW Unexpected firmware state.
Possible Causes
The Djinn Beetle generator reported a Heater 1 SW fault: state SW Unexpected firmware state.
Troubleshooting Steps
Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check