top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0204

Error Code System Message

​

0x0204, 0X0204, Brivo XR285amx 0x0204, Djinn Heater 2 error, Demand too low (target
Question

What does 0x0204 mean on a GE Brivo XR285amx?

Short Answer

0x0204: Demand too low (target < min heater).

Possible Causes

The Djinn Beetle generator reported a Heater 2 FW fault: Demand too low (target < min heater).

Troubleshooting Steps

Run Fil 2 Drv Clsd Loop, Fil 2 Xfmr Primary, and Fil 2 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0205

Error Code System Message

​

0x0205, 0X0205, Brivo XR285amx 0x0205, Djinn Heater 2 error, Inverter OFF and current not zero
Question

What does 0x0205 mean on a GE Brivo XR285amx?

Short Answer

0x0205: Inverter OFF and current not zero.

Possible Causes

The Djinn Beetle generator reported a Heater 2 FW fault: Inverter OFF and current not zero.

Troubleshooting Steps

Run Fil 2 Drv Clsd Loop, Fil 2 Xfmr Primary, and Fil 2 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0207

Error Code System Message

​

0x0207, 0X0207, Brivo XR285amx 0x0207, Djinn Heater 2 error, Over current detected by FPGA
Question

What does 0x0207 mean on a GE Brivo XR285amx?

Short Answer

0x0207: Over current detected by FPGA.

Possible Causes

The Djinn Beetle generator reported a Heater 2 FW fault: Over current detected by FPGA.

Troubleshooting Steps

Run Fil 2 Drv Clsd Loop, Fil 2 Xfmr Primary, and Fil 2 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0208

Error Code System Message

​

0x0208, 0X0208, Brivo XR285amx 0x0208, Djinn Heater 2 error, Filament 2 driver stopped by 48V out of range
Question

What does 0x0208 mean on a GE Brivo XR285amx?

Short Answer

0x0208: Filament 2 driver stopped by 48V out of range.

Possible Causes

The Djinn Beetle generator reported a Heater 2 FW fault: Filament 2 driver stopped by 48V out of range.

Troubleshooting Steps

Run Fil 2 Drv Clsd Loop, Fil 2 Xfmr Primary, and Fil 2 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0281

Error Code System Message

​

0x0281, 0X0281, Brivo XR285amx 0x0281, Djinn Heater 2 error, state SW Unexpected firmware state
Question

What does 0x0281 mean on a GE Brivo XR285amx?

Short Answer

0x0281: state SW Unexpected firmware state.

Possible Causes

The Djinn Beetle generator reported a Heater 2 SW fault: state SW Unexpected firmware state.

Troubleshooting Steps

Run Fil 2 Drv Clsd Loop, Fil 2 Xfmr Primary, and Fil 2 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page