Error 0x0283

Error Code System Message

​

0x0283, 0X0283, Brivo XR285amx 0x0283, Djinn Heater 2 error, Id SW Filament 2 driver diag failed
Question

What does 0x0283 mean on a GE Brivo XR285amx?

Short Answer

0x0283: Id SW Filament 2 driver diag failed.

Possible Causes

The Djinn Beetle generator reported a Heater 2 SW fault: Id SW Filament 2 driver diag failed.

Troubleshooting Steps

Run Fil 2 Drv Clsd Loop, Fil 2 Xfmr Primary, and Fil 2 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0400

Error Code System Message

​

0x0400, 0X0400, Brivo XR285amx 0x0400, Djinn Rotor error, Unkwn FW error
Question

What does 0x0400 mean on a GE Brivo XR285amx?

Short Answer

0x0400: Unkwn FW error.

Possible Causes

The Djinn Beetle generator reported a Rotor FW fault: Unkwn FW error.

Troubleshooting Steps

Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0401

Error Code System Message

​

0x0401, 0X0401, Brivo XR285amx 0x0401, Djinn Rotor error, CAN message received within 5 sec
Question

What does 0x0401 mean on a GE Brivo XR285amx?

Short Answer

0x0401: CAN message received within 5 sec.

Possible Causes

The Djinn Beetle generator reported a Rotor FW fault: CAN message received within 5 sec.

Troubleshooting Steps

Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0402

Error Code System Message

​

0x0402, 0X0402, Brivo XR285amx 0x0402, Djinn Rotor error, Database not correct
Question

What does 0x0402 mean on a GE Brivo XR285amx?

Short Answer

0x0402: Database not correct.

Possible Causes

The Djinn Beetle generator reported a Rotor FW fault: Database not correct.

Troubleshooting Steps

Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0403

Error Code System Message

​

0x0403, 0X0403, Brivo XR285amx 0x0403, Djinn Rotor error, Current overload
Question

What does 0x0403 mean on a GE Brivo XR285amx?

Short Answer

0x0403: Current overload.

Possible Causes

The Djinn Beetle generator reported a Rotor FW fault: Current overload.

Troubleshooting Steps

Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.