AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0x0404
Error Code System Message
Question
What does 0x0404 mean on a GE Brivo XR285amx?
Short Answer
0x0404: Open load.
Possible Causes
The Djinn Beetle generator reported a Rotor FW fault: Open load.
Troubleshooting Steps
Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0405
Error Code System Message
Question
What does 0x0405 mean on a GE Brivo XR285amx?
Short Answer
0x0405: Phases unbalanced.
Possible Causes
The Djinn Beetle generator reported a Rotor FW fault: Phases unbalanced.
Troubleshooting Steps
Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0406
Error Code System Message
Question
What does 0x0406 mean on a GE Brivo XR285amx?
Short Answer
0x0406: Phase error.
Possible Causes
The Djinn Beetle generator reported a Rotor FW fault: Phase error.
Troubleshooting Steps
Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0407
Error Code System Message
Question
What does 0x0407 mean on a GE Brivo XR285amx?
Short Answer
0x0407: Inverter permanent over current.
Possible Causes
The Djinn Beetle generator reported a Rotor FW fault: Inverter permanent over current.
Troubleshooting Steps
Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0408
Error Code System Message
Question
What does 0x0408 mean on a GE Brivo XR285amx?
Short Answer
0x0408: DC bus detected.
Possible Causes
The Djinn Beetle generator reported a Rotor FW fault: DC bus detected.
Troubleshooting Steps
Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.
Parts Areas to Check