Error Number

Error 0x0404

Error Code System Message

​

0x0404, 0X0404, Brivo XR285amx 0x0404, Djinn Rotor error, Open load
Question

What does 0x0404 mean on a GE Brivo XR285amx?

Short Answer

0x0404: Open load.

Possible Causes

The Djinn Beetle generator reported a Rotor FW fault: Open load.

Troubleshooting Steps

Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0405

Error Code System Message

​

0x0405, 0X0405, Brivo XR285amx 0x0405, Djinn Rotor error, Phases unbalanced
Question

What does 0x0405 mean on a GE Brivo XR285amx?

Short Answer

0x0405: Phases unbalanced.

Possible Causes

The Djinn Beetle generator reported a Rotor FW fault: Phases unbalanced.

Troubleshooting Steps

Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0406

Error Code System Message

​

0x0406, 0X0406, Brivo XR285amx 0x0406, Djinn Rotor error, Phase error
Question

What does 0x0406 mean on a GE Brivo XR285amx?

Short Answer

0x0406: Phase error.

Possible Causes

The Djinn Beetle generator reported a Rotor FW fault: Phase error.

Troubleshooting Steps

Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0407

Error Code System Message

​

0x0407, 0X0407, Brivo XR285amx 0x0407, Djinn Rotor error, Inverter permanent over current
Question

What does 0x0407 mean on a GE Brivo XR285amx?

Short Answer

0x0407: Inverter permanent over current.

Possible Causes

The Djinn Beetle generator reported a Rotor FW fault: Inverter permanent over current.

Troubleshooting Steps

Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0408

Error Code System Message

​

0x0408, 0X0408, Brivo XR285amx 0x0408, Djinn Rotor error, DC bus detected
Question

What does 0x0408 mean on a GE Brivo XR285amx?

Short Answer

0x0408: DC bus detected.

Possible Causes

The Djinn Beetle generator reported a Rotor FW fault: DC bus detected.

Troubleshooting Steps

Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.

Parts Areas to Check

​

For Related Models

Brivo XR285amx