AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0x0459
Error Code System Message
Question
What does 0x0459 mean on a GE Brivo XR285amx?
Short Answer
0x0459: Acceleration cmd while database not OK.
Possible Causes
The Djinn Beetle generator reported a Rotor FW fault: Acceleration cmd while database not OK.
Troubleshooting Steps
Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0460
Error Code System Message
Question
What does 0x0460 mean on a GE Brivo XR285amx?
Short Answer
0x0460: Database download while rotor speeding.
Possible Causes
The Djinn Beetle generator reported a Rotor FW fault: Database download while rotor speeding.
Troubleshooting Steps
Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0462
Error Code System Message
Question
What does 0x0462 mean on a GE Brivo XR285amx?
Short Answer
0x0462: Acceleration while in error.
Possible Causes
The Djinn Beetle generator reported a Rotor FW fault: Acceleration while in error.
Troubleshooting Steps
Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0464
Error Code System Message
Question
What does 0x0464 mean on a GE Brivo XR285amx?
Short Answer
0x0464: Speed value command not correct.
Possible Causes
The Djinn Beetle generator reported a Rotor FW fault: Speed value command not correct.
Troubleshooting Steps
Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0466
Error Code System Message
Question
What does 0x0466 mean on a GE Brivo XR285amx?
Short Answer
0x0466: 48V LVPS out of specification.
Possible Causes
The Djinn Beetle generator reported a Rotor FW fault: 48V LVPS out of specification.
Troubleshooting Steps
Run Rotation Test and Tube Rotor Impedance. Check rotor communication, tube rotation cable, stator resistance/isolation, Auxiliary LEDs/Neons, 48 V supply, and the error-data value for 0x0480 when applicable.
Parts Areas to Check