top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0504

Error Code System Message

​

0x0504, 0X0504, Brivo XR285amx 0x0504, Djinn BB error, DC bus rising time too slow
Question

What does 0x0504 mean on a GE Brivo XR285amx?

Short Answer

0x0504: DC bus rising time too slow.

Possible Causes

The Djinn Beetle generator reported a BB FW fault: DC bus rising time too slow.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0505

Error Code System Message

​

0x0505, 0X0505, Brivo XR285amx 0x0505, Djinn BB error, Buck Boost input current detected while ON
Question

What does 0x0505 mean on a GE Brivo XR285amx?

Short Answer

0x0505: Buck Boost input current detected while ON.

Possible Causes

The Djinn Beetle generator reported a BB FW fault: Buck Boost input current detected while ON.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0510

Error Code System Message

​

0x0510, 0X0510, Brivo XR285amx 0x0510, Djinn BB error, Over output DC voltage
Question

What does 0x0510 mean on a GE Brivo XR285amx?

Short Answer

0x0510: Over output DC voltage.

Possible Causes

The Djinn Beetle generator reported a BB FW fault: Over output DC voltage.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0511

Error Code System Message

​

0x0511, 0X0511, Brivo XR285amx 0x0511, Djinn BB error, Unknown BB state
Question

What does 0x0511 mean on a GE Brivo XR285amx?

Short Answer

0x0511: Unknown BB state.

Possible Causes

The Djinn Beetle generator reported a BB FW fault: Unknown BB state.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0520

Error Code System Message

​

0x0520, 0X0520, Brivo XR285amx 0x0520, Djinn BB error, Over input current
Question

What does 0x0520 mean on a GE Brivo XR285amx?

Short Answer

0x0520: Over input current.

Possible Causes

The Djinn Beetle generator reported a BB FW fault: Over input current.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page