AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0523

Error Code System Message

​

0x0523, 0X0523, Brivo XR285amx 0x0523, Djinn BB error, Zero Voltage Switching timeout
Question

What does 0x0523 mean on a GE Brivo XR285amx?

Short Answer

0x0523: Zero Voltage Switching timeout.

Possible Causes

The Djinn Beetle generator reported a BB FW fault: Zero Voltage Switching timeout.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0524

Error Code System Message

​

0x0524, 0X0524, Brivo XR285amx 0x0524, Djinn BB error, Buck-Boost over current
Question

What does 0x0524 mean on a GE Brivo XR285amx?

Short Answer

0x0524: Buck-Boost over current.

Possible Causes

The Djinn Beetle generator reported a BB FW fault: Buck-Boost over current.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0525

Error Code System Message

​

0x0525, 0X0525, Brivo XR285amx 0x0525, Djinn BB error, Boost timeout
Question

What does 0x0525 mean on a GE Brivo XR285amx?

Short Answer

0x0525: Boost timeout.

Possible Causes

The Djinn Beetle generator reported a BB FW fault: Boost timeout.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0580

Error Code System Message

​

0x0580, 0X0580, Brivo XR285amx 0x0580, Djinn BB error, Timeout when waiting for DC bus
Question

What does 0x0580 mean on a GE Brivo XR285amx?

Short Answer

0x0580: Timeout when waiting for DC bus.

Possible Causes

The Djinn Beetle generator reported a BB SW fault: Timeout when waiting for DC bus.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0581

Error Code System Message

​

0x0581, 0X0581, Brivo XR285amx 0x0581, Djinn BB error, input voltage detected too low
Question

What does 0x0581 mean on a GE Brivo XR285amx?

Short Answer

0x0581: input voltage detected too low.

Possible Causes

The Djinn Beetle generator reported a BB SW fault: input voltage detected too low.

Troubleshooting Steps

Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.

Parts Areas to Check

​

For Related Models

Brivo XR285amx