AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0x060A
Error Code System Message
Question
What does 0x060A mean on a GE Brivo XR285amx?
Short Answer
0x060A: kV zero.
Possible Causes
The Djinn Beetle generator reported a kV FW fault: kV zero.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x060C
Error Code System Message
Question
What does 0x060C mean on a GE Brivo XR285amx?
Short Answer
0x060C: Unknown kV regulation state.
Possible Causes
The Djinn Beetle generator reported a kV FW fault: Unknown kV regulation state.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x060E
Error Code System Message
Question
What does 0x060E mean on a GE Brivo XR285amx?
Short Answer
0x060E: bus voltage measured when error.
Possible Causes
The Djinn Beetle generator reported a kV FW fault: bus voltage measured when error.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x060F
Error Code System Message
Question
What does 0x060F mean on a GE Brivo XR285amx?
Short Answer
0x060F: Watchdog firmware error.
Possible Causes
The Djinn Beetle generator reported a kV FW fault: Watchdog firmware error.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0611
Error Code System Message
Question
What does 0x0611 mean on a GE Brivo XR285amx?
Short Answer
0x0611: RTL 2 error.
Possible Causes
The Djinn Beetle generator reported a kV FW fault: RTL 2 error.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.
Parts Areas to Check